| Wavelength Range | 190 to 900 nm |
| Spectral Bandwidth | 0.1, 0.2, 0.4, 1.0 and 2.0 nm |
| Wavelength Accuracy | ±0.15 nm |
| Wavelength Receptivity | ≤0.04 nm |
| Baseline Stability | ≤0.002A/30 min (Cu) |
| Characteristic Viscosity | 0.02 μg/ml/1% (Cu) |
| Detection Limit | 0.004 μg/ml (Cu) |
| Detection Accuracy | 0.005 |
| Grating | 1800 lines/ mm |
| Atomizer | Effective glass atomizer |
| Lamp Stand | 8 |
| D2 Background Correction Ability | When the background absorbance is 1 A, the system can reduce it by at least 50 times using a self-absorption background deduction method. |
| Burner Material | All-metal titanium burner |
| Burner Type | Acetylene air burner |
| Burner size | 100 mm |
| Ignition Baseline Drift | ≤0.006 A/30 min |
| Characteristic Viscosity (Cu) | ≤0.025 μg/ml/1% |
| Relative Standard deviation of Accuracy | ≤0.5% (Cu, absorbance>0.8A) |
| Detection Limit | Cu≤0.008 μg/ml |
| Safety system | Automatically cuts off gas in case of insufficient pressure, power outage, flameout, or burner misalignment |
| Graphite Furnace Maximum Temperature | 3000 ℃ |
| Maximum Temperature Rising Rate | ≥2000 ℃/s |
| Characteristic Detection Limit | Cd ≤ 0.5 × 10⁻¹² g, Cu ≤ 0.5 × 10⁻¹¹ g |
| Measurement Accuracy | Cu ≤ 3%, Cd ≤ 3% |
| Graphite Furnace Dimension (L×W×H) | 550 × 450 × 300 mm |
| Graphite Furnace Weight | 65 kg |
| Power Consumption | 7000 W |
| Power Supply | 220V ±22V AC |
| Size (L×W×H) | 700 × 550 × 450 mm |
| Weight | 75 kg |
Automatically adjustable position
Printable results
Fully Automated
Safety alarms present
Adjustable gas flow
Advanced temperature controls
Our Atomic Absorption Spectrophotometer BZAAS-502 is utilized in clinical, metallurgy, food, pharmaceutical and environmental testing sectors for trace metal detection and measuring the concentration of samples.