Micro Spectrophotometer BZPS-504 works as a new type of full-wavelength spectrophotometer. With the mode of USB output or network transfer data, with an electronic manual. Delivers a wavelength range of 190 to 850nm with a precision of ±1 nm. It uses a liquid column to recognize the irregular function. It results in a light source that is a long-life pulsed xenon flash lamp.
Wavelength Range | 190 to 850 nm |
---|---|
Wavelength Accuracy | ± 1 nm |
Wavelength Resolution | 2 nm (FWHM at Hg 546 nm) |
Repeatability | <± 0.5 nm |
Light Path | 1 mm, 0.05 mm (optical path automatic conversion) |
Nucleic Acid measurement range | 0.2 to 37500 mg / μl (sa DNA) |
Protein measurement range | 0.01 to 1120 mg / ml (BSA) |
sample volume requirements | 0.3 to 2 μl |
Absorbance Accuracy | 1% (0.76 absorbance at 350 nm) |
Absorbance Range | 0.002 to 300 Abs (equivalent to 10mm path length) |
Sample measurement time | less than 3s |
Sample base material | stainless steel and quartz fiber |
Light Source | long life pulsed xenon flash lamp |
Detector | 2048(CMOS) element linear silicide CCD array |
Users can accept measurement result data and custom methods directly in spreadsheet mode
32GB of storage space is pre-installed
Integrated fibre to prevent fibre breakage due to external force collision, which results in hectic measurement results
There is no need to warm up, start testing whenever you want
Built-in Win10 system, built-in 7-inch high-definition display, full touch operation
Stainless steel and quartz fibre used for the platform
Using a liquid column to identify abnormal function
The built-in Bluetooth and WIFI allow for remote operation and wireless printing of results
Two USB ports can be connected to a variety of devices
Nucleic acid, protein, bacteria, cell bacteria, microarray samples, and conventional full-wavelength scanning, among other things, can be detected.