Biological Safety Cabinet BZBN-505
Product1
Product2
Product3

Biological Safety Cabinet BZBN-505

Labzee Biological Safety Cabinet BZBN-505 designed for safe handling of biohazardous materials, supporting two-user operators. Features HEPA filtration, with an optional ULPA filter ensuring advanced air purification. Its consistent airflow at 0.55 m/s inflow and 0.35 m/s downflow ensures safe containment across 80 to 105 Pa. UV sterilization for automated decontamination, front window protection, and filter failure alarms for complete operational security.

$ 6775

Applicable Number Of Users 2
Cleanliness Class 100
Inflow Air Velocity 0.55 m/s
Down Airflow Velocity 0.35 m/s
Noise ≤ 67 dB
Vibrational Half Peak Value ≤ 5 µm
Illuminance ≥ 650 Lux
Ambient Temperature 10 to 30°C
Relative Humidity at 25°C ≤ 70 %
Atmospheric Pressure 80 to 105 Pa
Height of Rack 630 mm
Power 2000 W
Power And Number Of Lamp 30 W, 1 Piece
Power And Number Of UV Light 30 W, 1 Piece
Net Weight 243 Kg
Gross Weight 275 Kg
Internal Dimension 1360 × 580 × 595 mm
Housing Dimension 1500 × 770 × 2320 mm
Inflow HEPA Dimension 1395 × 475 × 70 mm
Exhaust HEPA Dimension 650 × 460 × 70 mm
Packaging Dimension 1760 × 970 × 1880 mm

LCD Real-Time Monitoring: Continuous display of airflow speed, filter status, and alerts.

Precision Airflow: Inflow 0.55 m/s and downflow 0.35 m/s for safe containment. 

UV Sterilization: Automated decontamination cycles for surface disinfection. 

Filter Failure Alarms: Preset alerts for filter damage or abnormal airflow. 

Front Window Protection: Safety interlock prevents operation with open window.

• Biohazardous Material Handling: Safe containment for infectious or toxic biological agents.
• Pharmaceutical Research: Contamination-free preparation of sterile drug compounds.
• Clinical Laboratories: Safe processing of patient samples and infectious specimens. 
• Research Facilities: Containment for cell culture, virology, and molecular biology. 
• Academic Institutions: Safe teaching environment for microbiology and biology students.

Frequently Asked Questions

1: Is the airflow monitored automatically in BZBN-505?

Yes, the intelligent air clean control system of BZBN-505 features a real-time LCD display that continuously monitors and displays inflow/downflow airspeed, filter life, fan/UV/light status, and alerts in case of abnormalities.

2: How does the BZBN-505 ensures user comfort during extended use?

BZBN-505 features a 10° inclined front window to reduce glare and wrist strain. The bright 650 Lux lighting and low vibration (<5 µm) provide a comfortable and stable work environment.

3: Is the BZBN-505 suitable for environments with fluctuating room conditions?

Yes, the BZBN-506 performs reliably within a temperature range of 10–30°C, ?70% humidity, and 80–105 Pa atmospheric pressure, making it ideal for most laboratory settings.

4: Does the BZBN-505 support compliance with international lab standards?

Yes, BZBN-505 uses ISO Class 5 HEPA filters, with optional ISO Class 4 ULPA filters, meeting high-level biosafety and cleanliness requirements for pharmaceutical, clinical, and research labs.

5: How does the BZBN-505 prevent contamination during power failure?

In case of power loss, the BZBN-505 ensures airtight sealing and passive protection of the work environment, while real-time data remains visible upon reboot.

6: What is the main purpose of a biological safety cabinet?

A biological safety cabinet is designed to protect the user, the environment, and the samples from biohazardous or infectious agents through filtered airflow and containment.

7: What kind of filtration system does the cabinet use?

Most BSCs use HEPA filters (High-Efficiency Particulate Air) to trap 99.99% of particles ?0.3 microns.

8: Does the biological safety cabinet offer protection against cross-contamination between samples?

Yes, Class II and III BSCs are designed to minimize cross-contamination with uniform laminar airflow and front barrier protection.


Biological Safety Cabinet BZBN-505

Biological Safety Cabinet BZBN-505