$ 5605
| Chamber Volume | 175 L |
| Temperature Range With Light | 10 to 60℃ |
| Temperature Range Without Light | 4 to 60℃ |
| Temperature Fluctuation | ±0.5℃ |
| Temperature Uniformity | ±2℃ |
| Operating Temperature | 5 to 30℃ |
| Humidity Range | 30 to 95% RH |
| Humidity Stability | ±3%RH |
| Light Intensity (LX) | 0 to 6000 (step control) |
| Timing Range | 1 to 99 Hours / Period |
| Sensor | PT100 Capacitance Sensor |
| Temp. And Humidity Adjusting | Balance Temperature And Humidity Adjusting |
| Power Consumption | 1650W |
| Chamber Size (W×D×H) | 45×42×93 cm |
| Exterior Size (W×D×H) | 65×73×175 cm |
3 Grids (Expandable to 8)
Dual Auto-Switch Refrigeration
JAKEL Circulation Fan
Multi-Segment Program Control
Dual-Parameter Digital Display
Built-In Data Printer
PID-Based Thermal Regulation
Labzee Pharmaceutical Light Stability Chamber BZML-501 is ideal for pharmaceutical stability testing, including accelerated aging, long-term storage, high humidity exposure, and photo stability assessments.
Frequently Asked Questions
1: How does Pharmaceutical Light Stability Chamber BZML-501 maintain stable conditions?
The PID microprocessor and PT100 sensor work with dual refrigeration systems for accuracy. The JAKEL fan and optimized airflow ensure ±2°C uniformity.
2: Why is Pharmaceutical Light Stability Chamber BZML-501 ideal for photo-stability testing?
Pharmaceutical Light Stability Chamber BZML-501 offers adjustable 0-6000 LX lighting with separate temperature controls (10-60°C with light, 4-60°C without). The UV sterilization and glass door enable contamination-free observation.
3: What's special about the airflow design?
The JAKEL fan and optimized ducts create uniform conditions (±2°C) throughout the 175L chamber. This prevents hot/cold spots even at full capacity.
4: What safety features protect samples?
Multiple alarms and automatic shutdown trigger if parameters exceed limits. Backup systems (mechanical thermostat, compressor protection) ensure failsafe operation.
5: Is Pharmaceutical Light Stability Chamber BZML-501 easy to use daily?
Pharmaceutical Light Stability Chamber BZML-501 features a menu-type operation interface on its large LCD screen that simplifies complex programming. The polished stainless-steel interior and external water tank make cleaning and maintenance straightforward.
6: What are Pharmaceutical Light Stability Chambers used for?
Pharmaceutical Light Stability Chambers are used to test the photostability of pharmaceutical products, packaging materials, and raw ingredients under controlled light conditions.
7: What type of lighting is used inside the chambers?
Our Pharmaceutical Light Stability Chambers use UV and visible fluorescent lamps with programmable light intensity, often ranging from 0 to 6000 Lux, to simulate daylight exposure.
8: Can temperature and humidity be controlled during testing?
Yes. Our chambers provide precise temperature and humidity control, with tight tolerances (e.g., ±0.5°C and ±3% RH), even under illuminated conditions.
9: How is light intensity monitored and controlled?
Built-in light sensors continuously monitor UV and visible light levels. The system allows step-controlled or programmable light exposure for reproducibility and compliance.
10: What safety features are included in these chambers?
Safety features include temperature-limiting alarms, UV protection shields, auto-shutdown mechanisms, and system diagnostics.