$ 11878
| Chamber Volume | 800 L |
| Temperature Range | 0 to 60℃ |
| Temperature Fluctuation | ±0.5℃ |
| Temperature Uniformity | ±2℃ |
| Operating Temperature | 5 to 30℃ |
| Humidity Range | 30 to 95% RH |
| Humidity Stability | ±3%RH |
| Timing Range | 1 to 99 Hours / Period |
| Sensor | PT100 Capacitance Sensor |
| Temp. And Humidity Adjusting | Balance Temperature And Humidity Adjusting |
| Power Consumption | 4000W |
| Chamber Size (W×D×H) | 96.5×61×137 cm |
| Exterior Size (W×D×H) | 110×93×227 cm |
3 Grids (Expandable to 10)
Touchscreen Microcomputer Control
Temperature-Limiting Alarm System
Stainless-Steel Chamber
Built-In UV Sterilization
Double-Layer Door System
Integrated Thermal Printer
Dual Auto-Switch Refrigeration
Labzee Pharmaceutical Stability Chamber BZEC-511 is ideal for accelerated, long-term, high humidity, and photo-stability testing of drugs and new formulations, supporting pharmaceutical stability studies.
Frequently Asked Questions
1: What safety mechanisms are integrated into the system?
Pharmaceutical Stability Chamber BZEC-511 includes a thermal limit alarm system, leakage protection, and a backup mechanical thermostat to ensure safe operation even in the event of primary control failure.
2: How does the chamber support data recording and compliance?
The Pharmaceutical Stability Chamber BZEC-511 is equipped with a USB port for easy data export and a built-in printer to log temperature and humidity parameters in real time—facilitating audit-ready documentation.
3: How does the BZEC-511 ensure accurate humidity control over time?
The system uses a high-quality international-brand humidity sensor paired with an external water tank, which minimizes fluctuations and ensures consistent RH levels. Combined with PID-based regulation, this setup allows for precise humidity control even in prolonged or variable testing conditions.
4: What is the purpose of the double-door design in the BZEC-511?
The double-door system enhances both insulation and usability. The inner door, made of tempered glass, allows users to monitor samples without disturbing internal conditions, while the magnetically sealed outer door ensures an airtight environment, improving energy efficiency and condition stability.
5: Can the BZEC-511 maintain uniformity across the entire 800L chamber space?
Yes, Pharmaceutical Stability Chamber BZEC-511 features a JAKEL circulation fan and a specially engineered air duct system that ensures consistent air distribution throughout the chamber. This design maintains uniform temperature and humidity levels across all shelves, even when fully loaded.
6: What is a Labzee Pharmaceutical Stability Chamber used for?
Labzee Pharmaceutical Stability Chambers are used to create controlled environments for testing drug stability, shelf life, and storage under ICH-guided temperature and humidity conditions.
7: What temperature and humidity range do these chambers support?
Labzee chambers offer a temperature range of 0°C to 60°C and a humidity range typically from 30% to 95% RH, with precise uniformity and control.
8: What chamber sizes are available?
They are available in multiple capacities ranging from 175 liters to 1075 liters, ideal for small labs to high-volume industrial testing.
9: How is temperature and humidity maintained inside the chamber?
Each unit features a high-efficiency airflow system and JAKEL circulation fans that ensure uniform distribution of temperature and humidity throughout the chamber.
10: What safety features are included?
The chambers are equipped with over-temperature protection, audio-visual alarms, and data logging systems for secure and traceable operation.
11: What type of control system does these chambers use?
Labzee chambers feature advanced PID microprocessor controllers with touchscreen interfaces for precise programming and monitoring of test parameters.