Download Catalog
Temperature Humidity Chamber BZHC-604 is a precise LED control interface conservational test chamber. It covers a wide temperature range of -20°C to 150°C, and humidity levels range from 20% to 98 % RH. It is made of inner stainless steel and cold-rolled steel with a powder-coated surface and can carry 800 liters of working space. For steady-state stability testing, basic temperature cycling, and speeded-up testing, imitate both temperature and humidity test procedures.
Temperature Range | -20°C to 150°C |
---|---|
Capacity | 800 L |
Temperature Fluctuation | ±0.5˚C |
Temperature Uniformity | ±2.0˚C |
Temperature Deviation | ±2.0˚C |
Humidity Range | 20% to 98% RH |
Humidity Fluctuation | ±2.5%RH |
Humidity Deviation | ±3.0%RH when humidity >75%RH, ±5.0%RH when humidity ≤75%RH |
Heat up rate non –Linear | -20°C to 150°C 3 ˚C /min |
Cool down rate | + 20°C to -20°C 1 ˚C /min |
Noise level | ≤70db |
Control tolerance | Temperature - ±0.20 ˚C humidity : ±1.0%RH |
Interface | USB, RS-232 or RS-485 |
Ambient conditions |
Temperature: +5°C to +35°C Humidity: ≤85%RH Air pressure: 86kPa to 106kPa |
Power | 9 KW |
Electrical Supply | AC380 (1±10%)V(50±0.5)Hz |
Internal Dimensions (W×D×H) | 1000×1000×800 mm |
External Dimensions (W×D×H) | 1250×2000×1800 mm |
There is also a smart balanced temperature adjustable control mode
It is powder-coated and made of stainless steel on the inside cold-rolled steel on the outside
A true-color LED touch screen display is used. and a control method that includes a program and a setpoint operating mode
A silicon seal, expanding foam insulation, and fiberboard sheeting are included
The system includes a nichrome heating resistance wire heating system and a centrifugal blower airflow fan system
Having a refrigeration method that does not deplete the ozone layer
It can be edited with 120 different programs, each with a 100-segment editing capability
To perform cold heat and humidity test, cold resistance test, thermal cycling test, ultra-low-temperature test, high-temperature test, storage condition, and calibration function.